Ahmed M. Increasing of temperature and spatial resolution of medical thermal imager with microbolometer matrix

Українська версія

Thesis for the degree of Candidate of Sciences (CSc)

State registration number

0416U000220

Applicant for

Specialization

  • 05.11.07 - Оптичні прилади та системи

19-01-2016

Specialized Academic Board

Д26.002.18

Essay

The mathematical model that describes the process of converting the infrared radiation from the region of the patient to the physician and taking in acount the thermogrammy preception by a doctor is developed.Based on this model the design methods of medical thermal imagers, which allow to predict generalized characteristics such as spatial and temperature resolution and to matching the parameters of the infrared lens and MBM to improve efficiency thermodiagnostics are proposed. Get more accurate equation to calculate the minimum resolvable temperature difference for the medical thermal imager.